6, Switch on Processor Board is Set
: This error o ccurs when
a service adjustment menu password can not b e set because
the override switch on the pro cessor is set. Qualied service
technicians, refer to \ADJUSTMENTS," in the
Agilent Technologies
8360 B-Series Swept Signal Generator/8360 L-Series Swept CW
Generator Service Guide
for more information.
Universal SCPI Error
Messages
Error Messages From
0
499 To
0
400
These error messages indicate that the Output Queue Control of the
swept signal generator has detected a problem with the message
exchange proto col. This type of error sets the Query Error Bit (bit
2) in the Event Status Register. One of the following has o ccurred:
An attempt has b een made to read data from the Output Queue
when no output is present or is p ending.
Data in the Output Queue has been lost.
Events that generate Query Errors do not generate Command Errors,
Execution Errors, or Device-specic Errors.
0
440, Query UNTERMINATED after indenite res
0
430, Query DEADLOCKED
0
430, Query DEADLOCKED;Output Buer Full
0
420, Query UNTERMINATED
0
420, Query UNTERMINATED;Nothing ToSay
0
410, Query INTERRUPTED
Error Messages From
0
399 To
0
300
These error messages indicate that some device op erations did not
properly complete, p ossibly due to an abnormal hardw
are or rmware
condition. This type of error sets the Device-sp ecic Error (bit 3)
in the Event Status Register. Events that generate Device-sp ecic
Errors do not generate Command Errors, Execution Errors, or Query
Errors.
0
350, To o many errors and also
0
32768
0
330, Self-test failed
0
330, Self-test failed;Power-On Tests
0
313, Calibration memory lost;Defaulted
Error Messages From
0
299 To
0
200
These error messages indicate that an error has b een detected by
the swept signal generator's Execution Control Blo ck. An error of
this type sets the Execution Error Bit (bit 4) in the Event Status
Register. One of the following events has o ccurred:
A data element following a header was evaluated by the swept
signal generator as outside of its legal input range or is inconsisten
t
with the swept signal generator's capabilit y.
2a-6 Error Messages

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?