Rack Flange Kit for
Swept Signal
Generators with
Handles Attached
(Option 913)
Option 913 swept signal generators are supplied with rack anges
and the necessary hardware to install them on the swept signal
generator without removing the instrument handles. The following
table itemizes the parts in this kit.
Table 3-6.
Rack Flange Kit for Swept Signal Generators with
Handles Attached Contents
Quantity Description
2 Rack Mount Flanges
8 Screws
CAUTION
Ventilation Requirements:
When installing the instrumentina
cabinet, the convection into and out of the instrumentmust not b e
restricted. The ambient temp erature (outside the cabinet) must be
less than the maximum op erating temp erature of the instrumen
t
by4
C for every 100 watts dissipated in the cabinet. If the total
power dissipated in the cabinet is greater than 800 w
atts, then forced
convection must b e used.
Installation 3-15