OPTION NOT INSTALLED. The language selected and the
corresp onding rmware/hardware necessary to run that language
is not present in the swept signal generator. See Chapter 3 for
information on language selection.
DISPLAY IS NOT RESPONDING
: Can app ear on the front panel
emulator if the internal processor can not communicate with the
display prop erly. This error indicates a display failure or a display
connector problem.
DIVIDE BY ZERO EXCEPTION
: This can only b e caused by
an internal pro cessor error. Refer to Chapter 4 for instructions on
contacting a qualied service technician.
EEROM FAILED, LOST CAL
: This error indicates that the swept
signal generator has lost its calibration constants and may not meet
specications. Refer to Chapter 4 and follow the lo cal op erator's
check pro cedures. If you are a qualied service technician and this
failure o ccurs, read the Calibration Constants section in the
Agilent
Technologies 8360 B-Series Swept Signal Gener
ator/8360 L-Series
Swept CW Generator Service Guide
.
EEROM Failed !!
: This error will only o ccur if the service
adjustment menu is accessed. Specically, an attempt has b een made
to write to a test patch and EEROM failed to store the data.
ERROR: CALIBRATION FAILED !!
: This error will only o ccur if
the service adjustment menu is accessed. Sp ecically, an A14 sweep
ramp calibration has been attempted and failed. Run the sw
eep
ramp selftest. Refer to the \MENU MAPS" chapter in the
Agilent
Technologies 8360 B-Series Swept Signal Generator/8360 L-Series
Swept CW Generator Service Guide
.
ERROR - Must rst enter correction freq
: This error o ccurs when
a correction p oint does not have its corresp onding frequency entered
rst. Refer to \Creating and Applying the User Flatness Correction
Array" in Chapter 1.
ERROR: Must rst enter a List Frequency !!
: This error o ccurs
when a dwell or oset value does not have its corresp onding
frequency entered rst. Refer to \Creating and Applying the User
Flatness Correction Array" in Chapter 1.
ERROR: Power SearchFailed !!
: This error o ccurs when the ALC
is in the ALC search mo de and is unable to level to the desired
power level. Refer to Chapter 4 and follow the lo cal op erator's check
pro cedures.
ERROR: Start must be
<
Stop !!
: This error o ccurs in asso ciation
with the frequency list, auto ll, feature. If the start frequency
entered is greater than the stop frequency,you will see this error.
Correct byentering a start frequency less than the stop frequency
.
ERROR: Stop must be
>
Start !!
: This error o ccurs in asso ciation
with the frequency list, auto ll, feature. If the stop frequency
2a-2 Error Messages