EasyManuals Logo
Home>GE>Medical Equipment>Revolution CT

GE Revolution CT User Manual

GE Revolution CT
702 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #356 background imageLoading...
Page #356 background image
5 Auto Gating
Auto Gating is a feature designed to optimize ECG-gated acquisition settings based on an
assessment of patient heart rate and variability, primarily focused on exams of the coronary
arteries. A similar feature on Discovery CT750 HD is referred to as SnapShot Assist. The Auto
Gating workflow provides the capability to record heart rate statistics during a breath hold to
determine the acquisition settings best suited for the heart rate characteristics of the patient.
For example, at higher heart rates, end-systolic phases are usually preferred due to the lack of
a diastasis period in mid-diastole. Auto Gating automatically uses heart-rate specific acquisition
settings (called “profiles”) from predetermined lookup tables.
A patient's heart rate can change over the course of a cardiac examination and result in
suboptimal results if adjustments are not made prior to the diagnostic cardiac CT scan. A
patient's heart rate should be assessed throughout an exam to watch for heart rate changes.
Consider the following factors that can impact a patient's heart rate during Coronary CT
Angiography (CCTA). These factors should remain consistent between the practice breath hold
and the gated cardiac exam to ensure that the pre-scan heart rate assessment is accurate.
Breath Hold Technique — end inspiration, mid inspiration, suspended breathing, et certera
Time Delay — after the breath hold command
Contrast — total volume, injection rate
Beta Blockers — oral and or IV
Vasodilators — can lead to slight increase in heart rate
Patient aspects — sex, age, electrophysiological disease, cardiac function issues
5.1 Auto Gating Acquisition Settings
Auto Gating generates suggested scan settings using the selected profile based on the patient's
recorded heart rate and selected Auto Gating profile (GE Adult Coronary). It uses the patient's
recorded heart rate information to predict the heart rate behavior during a CCTA scan to
optimize the parameters for each patient. You can acquire a recording automatically during scan
acquisition (timing bolus, calcium scoring series, etc.) or manually during a practice breath hold
by using the
Record New HR
option in the
Acquisition Window and Heart Rate
widget.
Illustration 13: Acquisition Window and Heart Rate widget
An alert message will be given under the following conditions:
Revolution CT User Manual
Direction 5480385-1EN, Revision 1
356 5 Auto Gating

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Revolution CT and is the answer not in the manual?

GE Revolution CT Specifications

General IconGeneral
BrandGE
ModelRevolution CT
CategoryMedical Equipment
LanguageEnglish

Related product manuals