5 System Operational Modes
5.1 Overview
The system provides powerful data collection capability with following scan modes:
Scout
Axial
Cine
Helical
Cardiac Axial
5.2 Scout
Scout imaging is used for anatomical location in conjunction with scan and recon prescription, to
provide an anatomical cross-reference for axial images, and to provide quick feedback to the
user as to the anatomy scanned. Scout supports the following features:
Selectable kV (70, 80, 100, 120, 140 kV)
Tube current limited to 250 mA with large focal spot
Approximately 100 mm/sec table speed
Detector coverage: 5 mm
Scout orientation: 0, 90, 180, 270 degrees
5.3 Axial
Axial imaging features include:
All kV and mA stations available, dependent on generator and tube limitations.
Rotation speeds: 0.28, 0.35, 0.5, 1.0 second
Detector coverage: 5, 40, 80, 120, 140, 160 mm
Variable image thickness selections: 0.625, 0.625z overlapped, 1.25, 2.5, and 5.0 mm
Sample rates: 984 Hz-8571Hz
Revolution CT can acquire up to 512 axial slices per rotation
5.4 Cine
Cine imaging features include:
All kV and mA stations available, dependent on generator and tube limitations
Maximum scan time: 60 seconds
Rotation speeds: 0.28, 0.35, 0.5, 1.0 second
Revolution CT User Manual
Direction 5480385-1EN, Revision 1
648 5 System Operational Modes