12-1
12 Cine Review
After you press the [Freeze] key, the system allows you to review and edit the images prior to
the image frozen. This function is called as cine review. The magnified images can also be
reviewed after the [Freeze] key is pressed, and the operating method is the same. You can
change post processing maps, perform measurements, add comments and body marks on
the images being reviewed.
The system supports manual review as well as automatic review. The default setup is Manual
Cine, but you can switch between Auto Cine and Manual Cine.
In addition, the system supports the images reviewed along with ECG waveforms, if the
detection of ECG waveforms is performed.
CAUTION:
1 Cine Review images can be inadvertently combined
in-between separate patient scans. The Cine Review
memory must be cleared at the end of the current
patient and the onset of the next new patient by
selecting the [End Exam] key on the control panel.
2 Cine files stored in the system’s hard drive shall contain
patient information, to avoid the selection of an
incorrect image file and potential misdiagnosis.
12.1 Entering / Exiting Cine Review
To enter Cine Review:
z Press [Freeze] key to freeze an image, and the [Cine] key indicator lights on;
z The system automatically enters the manual cine status.
To exit Cine Review:
z Press [Freeze] key again, the system will return to image scanning and exit cine review.
z Press [Cine] key or [Esc] key, the images are still frozen but the system exits cine review.
12.2 Cine Review in the B or C Mode
12.2.1 Manual Cine Review
After the cine review is entered in the B or C mode, rolling the trackball will display the cine
images on the screen one by one. If you roll the trackball to the left, the review sequence is
reversed to the image-storing sequence, thus the images are displayed in descending order.
Whereas, if you roll the trackball to the right, the review sequence is the same as the
image-storing sequence, thus the images are displayed in ascending order. When you review
images until the first or the last frame, further rolling the trackball will display the last or first
frame.