6-1
6 Power ON / OFF
6.1 Powering ON the System
CAUTION:
To ensure safe and effective system operation, you must
perform daily maintenance and checks. If the system begins to
function improperly – immediately stop scanning. If the system
continues to function improperly – fully shut down the system
and contact Mindray Customer Service Department or sales
representative. If you use the system in a persistent improperly
functioning state – you may harm the patient or damage the
equipment.
Checking before Power ON
To check the system before the system is turned on:
No Check Item
1
The temperature, relative humidity and atmospheric pressure shall meet the
requirements of operating conditions.
2
There shall be no condensation.
3 There shall be no distortion, damage or dirt on the system and peripheral
devices.
If any dirt is found, cleaning shall be performed as defined in section “22.1.1
Cleaning the System”.
4
There shall be no loose screws on the monitor or control panel.
5
There shall be no cable damage (e.g. power cord). Maintaining secure connections
to the system at all times.
6
The transducers and transducer cables shall be free of damage or stains.
See the transducer operator’s manuals for transducer cleaning and
disinfection.
7
No miscellaneous odds and ends are allowed to be attached or affixed to the
control panel.
8 Check the power cord / power supply and all I / O ports. Ensure that all
connections are free from damage and remain clear of foreign object
blockages.
There shall be no obstacles around the system and its air vent.
9 Prior to each use of a transducer, you should perform transducer cleaning and
disinfection as per the transducer operator’s manual.
10
The overall scanning environment and field must be clean.