Remote Commands Reference
R&S
®
RTC1000
188User Manual 1335.7352.02 ─ 02
CHANnel<m>:DATA:XINCrement?
Returns the time difference between two adjacent samples of the indicated waveform.
The command is relevant for data conversion, if binary data format is defined (FORM
UINT,8|16|32).
Suffix:
<m>
.
Selects the input channel (1, 2).
Return values:
<Xincrement> Time in s.
Usage: Query only
CHANnel<m>:DATA:XORigin?
Returns the time of the first sample of the indicated waveform. The command is rele-
vant for data conversion, if binary data format is defined (FORM UINT,8|16|32).
Suffix:
<m>
.
Selects the input channel (1, 2).
Return values:
<Xorigin> Time in s.
Usage: Query only
CHANnel<m>:DATA:YINCrement?
Returns the voltage value per bit of the indicated waveform. The command is relevant
for data conversion, if binary data format is defined (FORM UINT,8|16|32).
Suffix:
<m>
.
Selects the input channel (1, 2).
Return values:
<Yincrement> Voltage in V.
Usage: Query only
CHANnel<m>:DATA:YORigin?
Returns the voltage value for binary value 0 of the indicated waveform. The command
is relevant for data conversion, if binary data format is defined (FORM UINT,8|16|32).
Suffix:
<m>
.
Selects the input channel (1, 2).
Return values:
<Yorigin> Voltage in V.
Usage: Query only
Acquisition and Setup

Table of Contents

Other manuals for R&S RTC1000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the R&S RTC1000 and is the answer not in the manual?

R&S RTC1000 Specifications

General IconGeneral
BrandR&S
ModelRTC1000
CategoryTest Equipment
LanguageEnglish

Related product manuals