EasyManuals Logo
Home>Agilent Technologies>Inverter>83620B

Agilent Technologies 83620B User Manual

Agilent Technologies 83620B
496 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #349 background imageLoading...
Page #349 background image
SCPI COMMAND SUMMARY
Sets and queries the ALC leveling source selection switch.
The *RST value is
INTernal
.
POWer:ALC[:STATe] ON|OFF|1|0
POWer:ALC[:STATe]?
Sets and queries the state switch of the ALC. The p ositions are :
ON|normal ALC op eration
OFF|open lo op ALC mo de
When on, the power can be programmed in fundamental units as
selected by the
UNIT:POWer
command.
When o, the p ower is no longer calibrated in absolute units and is
set in units of dB of arbitrary mo dulator setting.
POWer:AMPLifier:STATE ON|OFF|0|1
POWer:AMPLifier:STATE?
Sets and queries the state of the amplier contained in
the doubler (for those mo dels with a doubler installed).
Programming a specic value for
POWer:AMPLifier:STATE
sets
POWer:AMPLifier:STATE:AUTO to OFF.
POWer:AMPLifier:STATE:AUTO ON|OFF|0|1
POWer:AMPLifier:STATE:AUTO?
Sets and queries the automatic selection of the doubler amplier
state. Programming a specic value for
POWer:AMPLifier:STATE
sets
POWer:AMPLifier:STATE:AUTO to OFF. *RST value is ON.
POWer:ATTenuation <num>[DB]|MAXimum|MINimum|UP|DOWN
POWer:ATTenuation? [MAXimum|MINimum]
Sets and queries the output attenuation level. Note that when
setting the attenuator level to 10 dB, the output p o
wer is
decreased by 10 dB. Programming a sp ecied attenuation sets
POWer:ATTenuation:AUTO OFF
.
POWer:ATTenuation:AUTO ON|OFF|1|0
POWer:ATTenuation:AUTO?
Sets and queries the state of the RF atten
uator coupling switch.
Programming a specied attenuation sets
POWer:ATTenuation:AUTO OFF
.
ON - insures that the amplitude level of the ALC is kept within
optimal limits.
OFF - the attenuator setting is set to the value of
POW:ATT
and left
there. The *RST value is
ON
.
POWer:CENTer <num> [lvl suffix]|MAXimum|MINimum|UP|DOWN
POWer:CENTer? [MAXimum|MINimum]
Sets and queries the center power for power sweep. Default units
(and units for query resp onse) are determined b
y the
UNIT:POWer
command.
S-42 Operating and Programming Reference

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 83620B and is the answer not in the manual?

Agilent Technologies 83620B Specifications

General IconGeneral
BrandAgilent Technologies
Model83620B
CategoryInverter
LanguageEnglish

Related product manuals