EasyManuals Logo
Home>Agilent Technologies>Inverter>83620B

Agilent Technologies 83620B User Manual

Agilent Technologies 83620B
496 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #352 background imageLoading...
Page #352 background image
SCPI COMMAND SUMMARY
POWer:STEP:AUTO ON|OFF|1|0
POWer:STEP:AUTO?
Sets and queries the function switch that controls how the power step
size (
POWer:STEP:INCRement
) is determined. If in the automatic
state, then the step size is 1 dB. The *RST setting is
ON
.
POWer:STEP[:INCRement] <num>[DB]|MAXimum|MINimum
POWer:STEP[:INCRement]? [MAXimum|MINimum]
Sets and queries the power step size to b e used for any node in the
power subsystem that allows
UP
and
DOWN
as parameters. Setting this
value explicitly causes
POWer:STEP:AUTO OFF
.
The *RST setting is 10 dB.
POWer:STOP <num>[lvl suffix]|MAXimum|MINimum|UP|DOWN
POWer:STOP? [MAXimum|MINimum]
Sets and queries the ending power for a p ower sweep. Default units
and units for query resp onse are determined by the command
UNIT:POWer
. The coupling equations for power sweep are exactly
analogous to those for frequency sweep. Power sweep is allowed to b e
negative, unlike frequency sweeps. *RST value is 0 dBm.
PULM:EXTernal:DELay <num>[time suffix]|MAXimum|MINimum
PULM:EXTernal:DELay? [MAXimum|MINimum]
Sets and queries the value of pulse delay from the time the external
pulse signal arrives to when the video pulse is generated. The
minimum value is 225 ns. After *RST the value is MINimum.
PULM:EXTernal:POLarity NORMal|INVerted
PULM:EXTernal:POLarity?
Selects the p olarity of the external pulse signal. NORMal causes
the positiv e-going edge of the signal to trigger the in
ternal pulse
generator and to turn on the RF. After *RST the value is NORMal.
PULM:INTernal:FREQuency <num>[freq
suffix]|MAXimum|MINimum
PULM:INTernal:FREQuency? [MAXimum|MINimum
Sets and queries the frequency of the in
ternal pulse generator.
The *RST value is 500 kHz.
PULM:INTernal:GATE ON|OFF|1|0
PULM:INTernal:GATE?
Sets and queries the state of the internal pulse generator's gating
control. When ON, and the pulse trigger source is internal, the
external pulse input is used to gate the pulse generator. When pulse
trigger source is external, this switch is ignored and no gating is
possible. After *RST, the setting is 0.
Operating and Programming Reference S-45

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 83620B and is the answer not in the manual?

Agilent Technologies 83620B Specifications

General IconGeneral
BrandAgilent Technologies
Model83620B
CategoryInverter
LanguageEnglish

Related product manuals