SCPI COMMAND SUMMARY
POWer:ALC:STATe POWer:SEARch Power Switch Action
ON ON ALC is momentarily closed to
level at the requested p ower,
and then the mo dulator is set to
the same voltage in op en lo op
mode. This rep eats automatically
any time that the p ower level or
frequency is changed.
ON OFF Normal mode.
ON ONCE Immediately p erforms a p ower
search. This leaves POWer:SEARch
in the ON p osition.
OFF not applicable Modulator setting is
explicitly set by user.
POWer:SLOPe <num>[DB/freq
suffix]|MAXimum|MINimum|UP|DOWN
POWer:SLOPe? [MAXimum|MINimum]
Sets and queries the RF slop e setting (dB p er Hz).
FREQ:MODE Aect on Slop e
CW or LIST Rotates around 0 Hz.
SWEep or
STEP
Rotates around the start frequency.
The *RST value is 0.
POWer:SLOPe:STATe ON|OFF|1|0
POWer:SLOPe:STATe?
Sets and queries the power slop e state. *RST value is 0.
POWer:SPAN <num>[DB]|MAXimum|MINimum|UP|DOWN
POWer:SPAN? [MAXimum|MINimum]
The coupling equations for p ower sweep are exactly analogous to
those for frequency sweep. Power sweep is allo wed to be negative,
unlike frequency sweeps. *RST value is 0.
POWer:STARt <num>[lvl suffix]|MAXimum|MINimum|UP|DOWN
POWer:STARt? [MAXimum|MINimum]
Default units and units for query resp onse are determined by the
command
UNIT:POWer
. The coupling equations for power sweep
are exactly analogous to those for frequency sweep. Power sweep is
allowed to b e negative, unlike frequency sweeps.
*RST value is 0 dBm.
POWer:STATe ON|OFF|1|0
POWer:STATe?
Sets and queries the output p ower on/o state. *RST value is
OFF
.
S-44 Operating and Programming Reference