Enter Correction Data into Array
12.
Select
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Mtr Meas Menu
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Measure Corr All
. The p ower meter
is now under swept signal generator control and is p erforming
the sequence of steps necessary to generate the correction
information at each frequency p oint.
If a GPIB error message is displayed verify that the interface
connections are correct. Check the GPIB address of the p ower
meter and ensure that it is the same address the swept signal
generator is using (address 13 is assumed). Refer to the menu
map 8, System, for the key sequence necessary to reach softkey
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Meter Adrs
.
Enable User Flatness Correction
13. When the op eration is complete, (a message is displayed) the
atness correction array is ready to be applied to your setup.
14. Tosave the swept signal generator parameters including the
correction table in an internal register, press
4
SAVE
54
n
5
.
(n = number 1 through 8).
15. Disconnect the p ower meter/sensor and press
4
FLTNESS ON/OFF
5
(amber LED on). The p ower pro duced at the p oint where the
power meter/sensor was disconnected is now calibrated at the
frequencies and power level sp ecied ab ove.
1-42 Getting Started Advanced

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 83620B and is the answer not in the manual?

Agilent Technologies 83620B Specifications

General IconGeneral
BrandAgilent Technologies
Model83620B
CategoryInverter
LanguageEnglish

Related product manuals