Note
U, V, and W-band p ower sensors are not available from
AgilentTechnologies. For these frequencies use the Anritsu ML83A
Power Meter with the MP715-004 (40 to 60 GHz), the MP716A
(50 to 75 GHz), or the MP81B (75 to 110 GHz) p ower sensors. Since
the Anritsu mo del ML83A Power Meter is not capable of internally
storing p ower sensor cal factors, you must manually correct the
data entry. Refer to example 2 for information on manual entry of
correction data.
Setup Swept Signal Generator Parameters
5. Turn on the swept signal generator and press
4
PRESET
5
.
The following occurs:
The source mo dule's frequency span is displayed on the swept
signal generator.
The swept signal generator's leveling mode is automatically
changed from internal to \mo dule lev
eling".
The source mo dule's maximum sp ecied p o
wer is set and
displayed.
6. Press FREQUENCY
4
START
54
2
54
6
54
.
54
5
54
GHz
5
,
4
STOP
54
4
54
0
5
4
GHz
5
. The frequency sweep is set from 26.5 to 40 GHz.
7. Press
4
POWER LEVEL
54
7
54
dBm
5
. The source mo dule power is set to
+7 dBm for maximum power to the device under test.
Access User Flatness Correction Menu
8. Press POWER
4
MENU
5
. Select
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Fltness Menu
.
9.
Select
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Delete Menu
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Delete All
. This step insures that the
atness array is empty.
10. Press
4
PRIOR
5
. Leave the delete menu and return to the previous
softkey menu.
11.
Select
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Enter Freq
4
2
54
6
54
.
54
5
54
GHz
5
,toenter 26.5 GHz as
the rst correction frequency.Enter 31, 32.5, and 40 GHz to
complete the list. Notice that the frequencies are arbitrarily
spaced.
Getting Started Advanced 1-41