3 DC Power Supply Operation
Protection Functions
82 U3606A User’s and Service Guide
Checking the OCP operation
To check OCP operation, raise the regulated output voltage slowly. Watch
the current drawn by the load as it reaches near the trip point. Then very
gradually increase the output voltage using the directional keys until the
OCP circuit trips. This will disable the U3606A output, cause the CV
annunciator to flash, and the OC and Error annunciator to illuminate.
The “triP” message appears on the display after a few seconds of
inactivity.
NOTE
The OCP protection feature is always enabled by default whenever you
select constant voltage mode. You cannot disable the OCP protection.
If the OCP value is set to a lesser value than the OC value, the OC value
will be adjusted to equal the OCP value.
The OCP value is not applicable for square-wave output.
The OCP value is limited by the range selected. Press Shift > Range to
select an appropriate range. You can only select a range when the
output is disabled (the SBY annunciator is illuminated).
NOTE
Error number 511, “Current output over protection” will be recorded in the
error queue when the OCP circuit trips. To read and clear the error
message, see “Reading error messages” on page 114. For a complete list
of all error messages, see Chapter 9, “List of Error Messages,” starting on
page 257.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies U3606A and is the answer not in the manual?

Agilent Technologies U3606A Specifications

General IconGeneral
BrandAgilent Technologies
ModelU3606A
CategoryMultimeter
LanguageEnglish

Related product manuals