Disassembly and Repair 7
Mechanical Disassembly
U3606A User’s and Service Guide 223
Mechanical Disassembly
For procedures in this section, the following tools are required for
disassembly:
T20 Torx driver (most disassembly)
#2 Pozi- drive screw driver
Flat blade screw driver
General disassembly
1 Turn off the power. Remove all cables from the instrument.
2 Remove the carry handle. Rotate the handle upright and pull out from
the sides of the instrument.
WARNING
SHOCK HAZARD. Only service-trained personnel who are aware of
the hazards involved should remove the instrument covers. To avoid
electrical shock and personal injury, make sure to disconnect the
power cord from the instrument before removing the covers. Some
circuits are active and have power applied even when the power
switch is turned off.