Characteristics and Specifications 8
Digital Multimeter Specifications
U3606A User’s and Service Guide 243
Duty cycle and pulse width specifications
Table 8 - 5 Frequency sensitivity for current measurement
Input range
Minimum sensitivity (rms sine wave)
20 Hz to 10 kHz
10 mA 1 mA
100 mA 10 mA
1.000 A 100 mA
3 A 300 mA
Table 8 - 6 Duty cycle and pulse width resolution and accuracy
Function Range Resolution Accuracy of full scale
Duty cycle 100.000%
[1]
[1] The range is calculated from {10 μs × frequency × 100%} to {[1 – (10 μs × frequency)] × 100%}. For example, a 1 kHz signal
can be measured from 1% to 99%.
0.001% 0.3% + 0.2% per kHz
Pulse width
199.999 ms
[2]
[2] The positive or negative pulse width must be greater than 10 μs. The range of the pulse width is determined by the frequency
of the signal.
0.001 ms Duty cycle/Frequency
1999.99 ms
[2]
0.01 ms Duty cycle/Frequency