8 Characteristics and Specifications
Digital Multimeter Specifications
238 U3606A User’s and Service Guide
Digital Multimeter Specifications
Specification assumptions
Specifications stated are for 60- minutes warm- up and 5½ digit
resolution.
One- year calibration cycle, with calibration temperature of 18 °C to
20 °C.
Operating temperature at 18 °C to 28 °C (64.4 °F to 82.4 °F).
Accuracy is expressed as: ±(% of reading + % of range)
Temperature coefficient: Add [0.1 × (the applicable accuracy) / °C] for
0 °C to 18 °C and 28 °C to 55 °C.
Relative humidity (RH) up to 80% at 30 °C, proportional to 50% for
30 °C to 55 °C.