2 Digital Multimeter Operation
Making Measurements
46 U3606A User’s and Service Guide
Performing capacitance measurements
The U3606A Multimeter|DC Power Supply calculates capacitance by
charging a capacitor with a known current for a period of time, and then
measuring the voltage.
Measuring tips:
For measuring capacitance values greater than 10000 μF, discharge the
capacitor first, then select a suitable range for measurement. This will
speed up the measurement time and also ensure that the correct
capacitance value is obtained.
For measuring capacitance values greater than 1 mF, a delayed
response should be expected from the front panel display. For remote
interface operation, increase the SCPI query timeout value. (Typically >
10000 ms.)
For measuring small capacitance values, press Null with the test leads
open to subtract the residual capacitance of the instrument and leads.
CAUTION
Disconnect circuit power and discharge all high-voltage capacitors
before measuring capacitance to avoid damaging the U3606A or the
device under test. To confirm that a capacitor has fully discharged, use
the DC voltage measurement. See “Measuring DC voltage” on page 26
for more information.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies U3606A and is the answer not in the manual?

Agilent Technologies U3606A Specifications

General IconGeneral
BrandAgilent Technologies
ModelU3606A
CategoryMultimeter
LanguageEnglish

Related product manuals