System Related Operation 4
Utility Menu Summary
U3606A User’s and Service Guide 111
Utility Menu Summary
The utility menu items are summarized in the table below. For further
explanations on the various items in the utility menu, see the respective
menu item headings.
Table 4 - 2 Utility menu item descriptions
Item Available settings Description
Error nonE (–)Er.NNN
• Review the last recorded error message (up to 20 error
messages).
• Review all the recorded error messages to clear the Error
annunciator.
• See “Reading error messages” on page 114 for more
information.
P.CodE NN.NN IOb NN.NN Sb NN.NN Mb
• Select “IOb” to view interface board program code revision.
• Select “Sb” to view source board program code revision.
• Select “Mb” to view measurement board program code
revision.
• See “Reading the program code revision” on page 115 for
more information.
diSP L-03 L-02 L-01
• Cycle through the VFD display brightness levels.
• See “Adjusting the display brightness” on page 116 for
more information.
P-on rESEt LASt
• Set “LASt” to recall the last known power-off state when
power is turned on.
• Set “rESEt” to recall the default factory power-on state
when power is turned on.
• See “Changing the power-on state” on page 116 for more
information.
bEEP 2400 Hz 3840 Hz OFF
• Set the beep driving frequency to “2400 Hz” or “3840 Hz”.
• Set “OFF” to disable the beep tone.
• See “Configuring the beeper” on page 117 for more
information.