Calibration Procedures 6
Calibration Message
U3606A User’s and Service Guide 183
Calibration Message
The instrument allows you to store a message in calibration memory. For
example, you can store information such as the date when the last
calibration was performed, the date when the next calibration is due, the
instrument serial number, or even the name and phone number of the
person to contact for a new calibration. The calibration message may
contain up to 40 characters.
You can record a calibration message only when the instrument is
unsecured.
You can read the calibration message (from the remote interface only)
whether the instrument is secured or unsecured.
To store a calibration message use the CALibration:STRing and
CALibration:STRing? commands from the remote interface.
See the U3606A Programmer’s Reference for more information on the
CALibration subsystem.
& CAL:STR "CAL: 27 Nov 2009"
This command stores a message in the
calibration memory.
& CAL:STR?
This query returns the message currently stored
in calibration memory (the quotes are also
returned).
$ "CAL: 27 Nov 2009"

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies U3606A and is the answer not in the manual?

Agilent Technologies U3606A Specifications

General IconGeneral
BrandAgilent Technologies
ModelU3606A
CategoryMultimeter
LanguageEnglish

Related product manuals