Page #173 background image
Verification and Performance Tests 5
Performance Verification Tests
U3606A User’s and Service Guide 151
Table 5 - 2 Zero offset verification test
Input Function Range
Error from
nominal 1 year
Short
DC voltage
100 mV ±0.008 mV
1 V ±0.00005 V
10 V ±0.0005 V
100 V ±0.005 V
1000 V ±0.05 V
Resistance
100 Ω ±0.008 Ω
1 kΩ ±0.00005 kΩ
10 kΩ ±0.0005 kΩ
100 kΩ ±0.005 kΩ
1 MΩ ±0.00005 MΩ
10 MΩ ±0.0005 MΩ
100 MΩ ±0.005 MΩ
Open DC current
10 mA ±0.0015 mA
100 mA ±0.005 mA
1 A ±0.00007 A
3 A ±0.00021 A