EasyManuals Logo
Home>Agilent Technologies>Multimeter>U3606A

Agilent Technologies U3606A User Manual

Agilent Technologies U3606A
288 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #52 background imageLoading...
Page #52 background image
2 Digital Multimeter Operation
Making Measurements
30 U3606A User’s and Service Guide
Measuring DC current
1 Press to make DC current measurement. (The DC annunciator is
shown on the display.)
2 Connect the red and black test leads to input terminals I (red) and
LO (black) respectively as shown in Figure 2- 2.
3 Probe the test points and read the display.
Table 2 - 3 DC current measurement summary
Item Description
Available ranges 10.0000 mA, 100.000 mA, 1.00000 A, 3.0000 A
Measurement method Sigma Delta A-to-D converter
Burden voltage and shunt
resistance
< 0.2 V, 10 Ω for 10 mA range
< 0.2 V, 1 Ω for 100 mA range
< 0.3 V, 0.1 Ω for 1 A range
< 0.7 V, 0.01 Ω for 3 A range
Input protection Protected with a 3.15 A/500 V, FF fuse
I
DC annunciator

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies U3606A and is the answer not in the manual?

Agilent Technologies U3606A Specifications

General IconGeneral
BrandAgilent Technologies
ModelU3606A
CategoryMultimeter
LanguageEnglish

Related product manuals