DC Power Supply Operation 3
Protection Functions
U3606A User’s and Service Guide 85
NOTE
The OV protection feature is always enabled by default whenever you
select constant current mode. You cannot disable the OV protection.
If the OV value is set to a greater value than the OVP value, the OVP
value will be adjusted to equal the OV value.
If the OV value is set to zero, this will result in the output current being
dropped to zero for limiting.
The OV value is not applicable for square-wave output.
The OV value is limited by the range selected. Press Shift > Range to
select an appropriate range. You can only select a range when the
output is disabled (the SBY annunciator is illuminated).