Page #270 background imageLoading...
Page #270 background image
8 Characteristics and Specifications
Digital Multimeter Specifications
248 U3606A User’s and Service Guide
Measurement noise Rejection
CMRR (common mode
rejection ratio) for 1 kΩ
unbalanced in LO lead
DC: 140 dB
AC: 70 dB
NMRR (Normal Mode
Rejection Ratio)
60 Hz ± 0.1%
5½ digits: 65 dB
4½ digits: 0 dB
50 Hz ± 0.1%
5½ digits: 55 dB
4½ digits: 0 dB