8 Characteristics and Specifications
DC Power Supply Specifications
254 U3606A User’s and Service Guide
[3] Specification applies when the positive or negative pulse width is greater than 50 μs.
[4] For frequency signals greater than 100 Hz, an additional 0.1% per 100 Hz is added. The accuracy of the pulse width should
be calculated as:
Calculation example: Frequency setting = 4800 Hz, Duty cycle setting = 50%
The pulse width accuracy (for frequency setting 4800 Hz and duty cycle setting 50%) is calculated as 0.2083 ms ±10.625 μs.
Accuracy 0.4%
frequency
100
----------------------------- 1
⎝⎠
⎛⎞
0.1%×+
⎝⎠
⎛⎞
=
Specification of pulse width Accuracy Pulse× width setting()±=
Specification of pulse width 0.4%
4800
100
------------ 1
⎝⎠
⎛⎞
0.1%×+
⎝⎠
⎛⎞
1
4800
------------
×
⎝⎠
⎛⎞
±
5.1
100
---------
1
4800
------------
×
⎝⎠
⎛⎞
± 10.625± μs===
NOTE
The rise and fall time is less than 25 μs.
The additional load regulation is 0.15 V/A.