2 Digital Multimeter Operation
Math Operations
58 U3606A User’s and Service Guide
Editing the null value
1 Press Null again to view and edit the null value.
2 Press or to select which digit position or range to be changed.
3 Press or to increase or decrease the value.
4 Press Shift > Save to save the changes. “donE” will be briefly displayed
in the upper secondary display. The display will then return to normal.
5 Press Shift > Exit to exit the edit mode without saving.
NOTE
In resistance measurement, the U3606A will read a non-zero value
even when the two test leads are in direct contact, because of the
resistance of these leads. Use the null function to zero-adjust the
display.
In DC voltage measurement mode, the thermal effect will influence the
accuracy. Short the test leads and press Null once the displayed value
is stable to zero-adjust the display.