6.3.2 Low Pressure Alarm
Property Description
Alarm text
Low Pressure
Priority
High
Alarm condition A Low Pressure alarm will be given when the Vivo 45 pres-
sure fails to reach the low pressure alarm limit for 15 seconds.
The low pressure alarm is disabled during ramp periods.
Possible cause
Disconnection of patient circuit.
Mismatch between pressure setting and alarm setting.
Leakage from the mask or other components of the
patient circuit.
Reset criteria
The pressure rises above the alarm limit.
Ventilator action The Vivo 45 will continue treatment according to the current
settings.
Setting range
1 cmH
2
O to 40 cmH
2
O
Note that the Low pressure alarm cannot be set higher than
the value set for Insp. Pressure or the High pressure alarm.
Setting resolution Below 10 cmH
2
O: 0.5 cmH
2
O
Above 10 cmH
2
O: 1.0 cmH
2
O
Setting display The Low Pressure alarm setting is displayed by a red line in
the pressure bar graph.
Alarms 103
Doc. 006149 D-3
Vivo 45 User manual

Table of Contents

Other manuals for Breas Vivo 45

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Breas Vivo 45 and is the answer not in the manual?

Breas Vivo 45 Specifications

General IconGeneral
BrandBreas
ModelVivo 45
CategoryMedical Equipment
LanguageEnglish

Related product manuals