6.4.20 Alarm Battery Error Alarm
Property Description
Alarm text
Alarm Battery Error
Alarm condition Unable to communicate with super capacitor and read super
capacitor status.
Priority
Medium
Ventilator action The ventilator will continue treatment with the same settings.
Reset
When triggering condition is removed.
6.4.21 Internal/Click-In Battery Hot Alarm
Property Description
Alarm text
Internal Battery — Internal Battery Hot
Click-In Battery — Click-In Battery Hot
Alarm condition
An alarm for Internal/Click-In Battery Overheat in Discharge
will be given when the internal or click-in battery reaches 55°
C (131°F).
NOTE
The battery electronics by manufacture stops dis-
charge at 60°C (140°F).
Priority High
Ventilator action The ventilator will continue treatment with the same settings.
Battery discharging will be disabled (by the battery elec-
tronics) once the temperature gets to 60°C (140°F). (If the
battery is last power source, the ventilator will stop running).
6.4.22 Heated Circuit Temp. Alarm
Property Description
Alarm text
Heated Circuit Temp.
Alarm condition A Heated Circuit temp alarm will be given when the meas-
ured temperature of the heated wire is outside the tolerance.
Priority
Medium
Ventilator action The ventilator will continue treatment with the same settings.
Priority
Medium
Reset
Heated wire measured temp tolerance is inside limits.
126 Alarms
Vivo 45 User manual
Doc. 006149 D-3