Unit Min Max Default
cmH
2
O 4 20 10
(A)
8
(P)
(A)= Adult mode, (P)= Paediatric mode
5.5.15 Ramp Time
The Ramp Time setting defines the time during which the pressure will increase from
the set Start Pressure to the set Inspiratory Pressure.
Ramp Time can only be set in Clinical Mode.
Unit Min Max Default
Minute 10
Off
60
Off
5.5.16 Start Pressure
The Ramp Start Pressure defines the pressure level delivered by the ventilator at the
beginning of the Ramp Time.
Unit Min Max Default
cmH
2
O 2 Current
Insp. Pressure
–2 or Current CPAP
5
5.5.17 Always Start Ramp
The Always Start Ramp determines whether the ramp is started automatically at the
beginning of the treatment.
Unit Min Max Default
—
Off
On
Off
5.5.18 Humidifier
Humidifier allows the user to start or stop the heated humidification.
The click-in water chamber needs to be connected before the setting can be turned On.
Unit Min Max Default
—
Off
On
Off
5.5.19 Humidifier Setting
The Humidifier Setting defines the level of humidity of the air delivered to the patient.
How to Use the Vivo 45 71
Doc. 006149 D-3
Vivo 45 User manual