6.3.3 High PEEP Alarm
Property Description
Alarm text
High PEEP
Priority
Medium
Alarm condition A High PEEP alarm will be given when the measured PEEP
is 30% above the set value for more than 15 seconds
Possible cause
Blocked leakage port.
Too short expiratory time.
Changes in airway resistance and or compliance.
Reset criteria
PEEP has gone below the alarm limit (lower than 30% above
the set value).
Ventilator action The Vivo 45 will continue treatment according to the current
settings.
Setting range
On
Off
6.3.4 Low PEEP Alarm
Property Description
Alarm text
Low PEEP
Priority
Medium
Alarm condition A Low PEEP alarm will be given when the measured PEEP
is 30% below the set value for more than 60 seconds
Possible cause
Excessive leakage.
Reset criteria
PEEP has gone above the alarm limit (higher than 30%
below the set value).
Ventilator action The Vivo 45 will continue treatment according to the current
settings.
Setting range
On
Off
104 Alarms
Vivo 45 User manual
Doc. 006149 D-3

Table of Contents

Other manuals for Breas Vivo 45

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Breas Vivo 45 and is the answer not in the manual?

Breas Vivo 45 Specifications

General IconGeneral
BrandBreas
ModelVivo 45
CategoryMedical Equipment
LanguageEnglish

Related product manuals