6.3.24 Low EtCO
2
Alarm
Property Description
Alarm text
Low EtCO2
Priority
Medium
Alarm condition A Low EtCO
2
alarm will be given when the measured EtCO
2
is below the alarm limit for 30 seconds.
Possible cause
Alarm limit is set too high.
Ventilator disconnection.
Excessive leakage in the Patient circuit/Interface.
Partial obstruction of the airways.
Breath Rate too high.
Delivered Tidal Volume too high.
Self triggering of the ventilator.
Setting range 1 to 99 mmHg
Off
Setting resolution
1 mmHg
Ventilator action The ventilator will continue treatment with the same settings.
This alarm requires a connected EtCO
2
sensor
6.3.25 High InspCO
2
Alarm (High Inspired CO
2
)
Property Description
Alarm text
High InspCO2
Priority High
Alarm condition A High Inspired CO
2
alarm will be given when the measured
inspired CO
2
exceeds the alarm limit for 30 seconds.
Possible cause
Alarm limit is set too low.
Excessive dead space between patient and leakage port.
Leakage port occluded.
Setting range 1 to 99 mmHg
Off
Setting resolution
1 mmHg
Ventilator action The ventilator will continue treatment with the same settings.
This alarm requires a connected EtCO
2
sensor
116 Alarms
Vivo 45 User manual
Doc. 006149 D-3

Table of Contents

Other manuals for Breas Vivo 45

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Breas Vivo 45 and is the answer not in the manual?

Breas Vivo 45 Specifications

General IconGeneral
BrandBreas
ModelVivo 45
CategoryMedical Equipment
LanguageEnglish

Related product manuals