8.2.4 Monitored Values Specifications
This section describes the ranges and tolerances for monitored values on the Vivo 45.
All stated tolerances includes measurement uncertainty. The accuracies have been tested
with all allowed configurations. Stated tolerances only disclose the maximum tolerance.
P
peak
Range/Performance: 4 to 99 cmH
2
O.
Resolution: ±0.5 cmH
2
O or ±10%, whichever is greatest
PEEP
Range/Performance: 0 to 99 cmH
2
O.
Resolution: ±0.5 cmH
2
O or ±10%, whichever is greatest
P
mean
Range/Performance: 0 to 99 cmH
2
O.
Resolution: ±0.5 cmH
2
O or ±10%, whichever is greatest
Leakage
Range/Performance: 0 to 99.9 l/min (BTPS*).
Resolution: ±10%
MV
i
Range/Performance: 0 to 99.9 l (BTPS*).
Resolution: ±10% or (±15 ml × bpm), whichever is greatest
MV
e
Range/Performance: 0 to 99.9 l (BTPS*).
Resolution: ±15% or (±15 ml × bpm), whichever is greatest
Vt
i
Range/Performance: 0 to 9999 ml (BTPS*).
Resolution: Exhalation valve circuit:±15 ml or 10%, whichever is greatest. Leakage cir-
cuit:±15 ml or 15%, whichever is greatest
148 Technical Specifications
Vivo 45 User manual
Doc. 006149 D-3

Table of Contents

Other manuals for Breas Vivo 45

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Breas Vivo 45 and is the answer not in the manual?

Breas Vivo 45 Specifications

General IconGeneral
BrandBreas
ModelVivo 45
CategoryMedical Equipment
LanguageEnglish

Related product manuals