Property Description
Alarm text
Disconnection
Priority
High
Alarm condition A Disconnection alarm will be given when the measured
flow exceeds the expected leakage flow at the set Pressure for
15 seconds.
Possible cause
Too high leakage in the patient circuit.
The patient has removed the mask.
Circuit disconnection.
Pilot pressure tube disconnection
Reset criteria
The leakage is back within limits.
Ventilator action The Vivo 45 will continue treatment according to the current
settings
Setting range
On
Off
6.3.17 Rebreathing Alarm
Property Description
Alarm text
Rebreathing
Priority
High
Alarm condition Leakage Circuit
A Rebreathing alarm will be given if the intentional leakage is
too low for more than 15 seconds.
MPV circuit
A Rebreathing alarm will be given if air returns into the venti-
lator for more than 10 consecutive breaths.
Possible cause
Obstructed or occluded patient circuit.
Incorrect patient circuit.
Obstructed or removed CO
2
port from leakage circuit.
Reset criteria
The leakage is back within limits.
Ventilator action The Vivo 45 will continue treatment according to the current
settings.
Setting range
On
Off
112 Alarms
Vivo 45 User manual
Doc. 006149 D-3

Table of Contents

Other manuals for Breas Vivo 45

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Breas Vivo 45 and is the answer not in the manual?

Breas Vivo 45 Specifications

General IconGeneral
BrandBreas
ModelVivo 45
CategoryMedical Equipment
LanguageEnglish

Related product manuals