EasyManuals Logo

Breas Vivo 45 User Manual

Breas Vivo 45
184 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #120 background imageLoading...
Page #120 background image
6.4.3 Low Patient Air Temp. (Low Patient Air Temperature Alarm)
Property Description
Alarm text
Low Patient Air Temp.
Alarm condition A Low Patient Air Temperature alarm will be given when the
patient air temperature is below the preset limit -30°C (-22°F).
Priority High
Possible cause Too low ambient temperature
Ventilator action The ventilator will continue treatment with the same settings.
6.4.4 Low Last Power Source Alarm
Property Description
Alarm text
Low Last Power Source
Priority
Medium
Alarm condition This alarm will be given when the last battery source (internal
battery) has 15 minutes of operating time left with current
settings.
Ventilator action The Vivo 45 will continue treatment according to the current
settings.
6.4.5 Crit. Low Last Power Source Alarm
Property Description
Alarm text
Crit. Low Last Power Source
Alarm condition A Crit. Low Last Power Source alarm will be given when the
last battery source (internal battery or click-in battery) has
5 minutes of operating time left with current settings.
Priority High
Ventilator action The ventilator will continue treatment with the same settings.
Reset
Connection of higher power source.
120 Alarms
Vivo 45 User manual
Doc. 006149 D-3

Table of Contents

Other manuals for Breas Vivo 45

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Breas Vivo 45 and is the answer not in the manual?

Breas Vivo 45 Specifications

General IconGeneral
BrandBreas
ModelVivo 45
CategoryMedical Equipment
LanguageEnglish

Related product manuals