Ingress Protection:
IP22
Solid particle protection: Hazardous parts are protected from touch by fingers and by
objects greater than 12 mm.
Liquid ingress protection: The protection withstands dripping water less than 15 degrees
from vertical.
The ingress protection has been tested by water drips equivalent to 3mm rain/minute
for 10 minutes (2.5 minutes for each tilting direction).
8.2.7 Other
Patient Circuit Leakage
Recommended leakage: 20 to 50 l/min at 10 cmH
2
O (leakage circuit)
Minimum leakage: 12 l/min at 4 cmH
2
O (leakage circuit)
Oxygen Inlet
Oxygen inlet port: Maximum flow: 15 l/min (medical oxygen). Oxygen coupling is
type CPC PMCD181032.
Start-up Time
Start-up from unpowered state: about 20 seconds.
Sound Power Level
Sound level at 10 cmH
2
O in CPAP mode: Less than 30 dB(A). Measured at 1 m.
Alarm sound level : Adjustable 5080 dB(A), Measured at 1m. Tolerance: ± 5 dB(A).
Miscellaneous
Maximum flow: > 300 l/min
Maximum flow at 20 mbar: > 150 l/min
Maximum limited pressure during single fault condition: 80 cmH2O (PCV, PSV)
30 cmH2O (CPAP)
Breathing resistance under single-fault: <6 cmH2O at 30 l/min, <6 cmH2O at 60
l/min
Vivo 45 Dimensions
W × H × D: 216 × 159 × 152 mm
Weight: 2.4 kg
Patient air outlet: 22 mm male, conical standard connector
Technical Specifications 151
Doc. 006149 D-3
Vivo 45 User manual

Table of Contents

Other manuals for Breas Vivo 45

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Breas Vivo 45 and is the answer not in the manual?

Breas Vivo 45 Specifications

General IconGeneral
BrandBreas
ModelVivo 45
CategoryMedical Equipment
LanguageEnglish

Related product manuals