*: These ventilator settings affect the operating time significantly.
Battery
Operating Time
Internal Battery 2.5 h
Click-in Battery 6.5 h
5.6.8 Storing the Internal Battery and the Click-in Battery
Storage longer than 1 month should be initiated with half-charged batteries in order to
maintain maximum capacity. Optimal storage temperature is 5 to 30°C (41 to 86°F).
5.6.9 External DC
WARNING!
Do not connect the Vivo 45 to a wheelchair unless the operating manual for the
wheelchair permits this as this can affect the Vivo 45 performance and consequently
result in patient death.
The ventilator can be operated from:
A 12 V external DC source using the 12/24 V converter.
A 24 V external DC source using the external battery cable.
Both AC power supply and external DC using the Y-cable.
With both AC power supply and external DC source connected, the ventilator will auto-
matically switch over to the external DC source if the AC power supply fails. The exter-
nal DC voltage level is shown under Others, Device Information in the menu
1 Connect the external DC cable to
the ventilator. Make sure that it is
fitted correctly.
2 Connect the other end of the cable to the DC source.
78 How to Use the Vivo 45
Vivo 45 User manual
Doc. 006149 D-3

Table of Contents

Other manuals for Breas Vivo 45

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Breas Vivo 45 and is the answer not in the manual?

Breas Vivo 45 Specifications

General IconGeneral
BrandBreas
ModelVivo 45
CategoryMedical Equipment
LanguageEnglish

Related product manuals