5.7.2.1 Connect the ventilator to a Nurse Call System
1 Connect the nurse call cable at the
back of the ventilator.
2 Test the connection by triggering an alarm on the ventilator and verify that the
nurse call system activates.
5.7.3 Using the ventilator with the FiO
2
Sensor
The FiO
2
sensor can be used to monitor and store FiO
2
measurements. The FiO
2
sen-
sor measures the fraction of inspired oxygen in the air channel of the ventilator. The
FiO
2
measurements will be stored in the data memory which can be downloaded to a
PC and viewed in Breas software.
Usage
Time
Operating temperature
10 to 40°C (50 to 104°F)
Operating pressure
700 to 1250 mbar
Expected operating life <3 years (in ambient air) or 500,000 Vol.% h,
whichever comes first.
Shelf life < 6 months (recommended)
CAUTION!
Note that the operating conditions for the FiO
2
sensor is different from the ventilator
system conditions. If the sensor is used outside its operating conditions the FiO
2
measurements might deviate.
5.7.3.1 Installing the FiO
2
Sensor
1 Place the ventilator so the bottom is accessible.
80 How to Use the Vivo 45
Vivo 45 User manual
Doc. 006149 D-3

Table of Contents

Other manuals for Breas Vivo 45

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Breas Vivo 45 and is the answer not in the manual?

Breas Vivo 45 Specifications

General IconGeneral
BrandBreas
ModelVivo 45
CategoryMedical Equipment
LanguageEnglish

Related product manuals