WARNING!
Do not use the device and contact your responsible care provider for an inspection
of the device in the event of:
• Unexpected patient symptoms during treatment.
• Unexplainable or sudden pressure, performance or sound changes during
operation.
• Suspected damage to the device, including the occurrence of Internal Functional
Failure alarms.
• Suspected damage to the click-in battery, including evidence of battery cell leakage.
Authorised service workshops can order the ventilator Service Manual that contains
all technical documentation required for the maintenance and service of the ventilator.
7.5 Storage
Store the ventilator in a dark room, where the temperature range is within -20 to +60°C
(-4 to +140°F).
CAUTION!
The ventilator must not be stored in a warm place, such as direct sunlight or close to
a radiator. The time required for the device to cool from the maximum storage tem-
perature of +60°C (+140°F) until it is ready for use in ambient temperature of +20°C
(+68°F) is 30 minutes.
If stored in a cold environment, let the ventilator adapt to room temperature before
using the device. The time required for the device to warm from the minimum stor-
age temperature of -20°C (-4°F) until it is ready for use in ambient temperature of
+20°C (+68°F) is 30 minutes.
7.6 Disposal
The ventilator, any accessories and all replaced parts must be disposed of and recycled
in accordance with the local environmental regulations regarding the disposal of used
equipment and waste. Contact your service provider for information regarding the dis-
posal procedure.