Insp. Time
Range/Performance: 0.3 to 5 s.
Resolution: ±0.1 s
Rise Time
Range/Performance: 0.1 to 5 s.
Resolution: ±10% or ±0.1 s, whichever is greatest
EtCO
2
Range/Performance: 0 to 25%.
Resolution: 0 to 15%: ±(0.3 vol% + 4% of reading).15 to 25%: unspecified
InspCO
2
Range/Performance: 0 to 25%.
Resolution: 0 to 15%: ±(0.3 vol% + 4% of reading).15 to 25%: unspecified
8.2.5 Power Supply
AC supply: 100 to 240 V AC, tolerance: +10%/-20%, 50 to 60 Hz, 1.0 - 2.0 A.
External DC: 19 V DC, tolerance: 19 V ± 6 V. Max 90 W.
Click-in battery: Capacity: 65Wh. Li-ion.
Internal battery: Capacity: 25Wh. Li-ion. Expected service life: 500 full charging cycles.
8.2.6 Environmental Conditions
Operating temperature range: 5 to 40°C (41 to 104°F)
Storage and transport temperature: -20 to +60°C (-4 to +140°F)
Ambient pressure range:
700 to 1100 mbar, corresponding to
~4200 metres (13800 feet) above sea
level to ~700 metres(2300 feet) below
sea level, at normal atmospheric
pressure.
As seen in the graph above, the ventila-
tor is unable to deliver set max pressure
at a very low ambient pressure.
150 Technical Specifications
Vivo 45 User manual
Doc. 006149 D-3