6.4.28 Internal Error Alarm
Property Description
Alarm text
Internal Error
Priority High
Alarm Condition An internal Error alarm will be given when the ventilator has
an internal error, followed by an error code for the specific
failure. All Internal Error alarm error codes are defined and
explained in the ventilator Service Manual.
Ventilator action The ventilator will continue treatment with the same settings.
Reset action
Power off and restart the ventilator.
6.4.29 Database Integrity Fail Alarm
Property Description
Alarm text
Database Integrity Failed
Priority High
Alarm Condition This alarm is given when the database integrity check fails.
Ventilator action The ventilator will continue treatment with the same settings.
Reset action
Rebuild the database and restart the ventilator.
6.4.30 Cooling Fan Error Alarm
Property Description
Alarm text
Cooling Fan Error
Alarm Condition The Cooling Fan Error alarm shall be given when the cooling
fan runs too slow.
Priority High
Ventilator action The ventilator will continue treatment with the same settings.
Reset
When cooling fan speed is above 275 rpm.
Alarms 129
Doc. 006149 D-3
Vivo 45 User manual

Table of Contents

Other manuals for Breas Vivo 45

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Breas Vivo 45 and is the answer not in the manual?

Breas Vivo 45 Specifications

General IconGeneral
BrandBreas
ModelVivo 45
CategoryMedical Equipment
LanguageEnglish

Related product manuals