• The maximum load of the trolley basket is 0.9 kg (2 lbs).
• The maximum load of the IV-pole is 3 kg (6.5 lbs).
• The maximum load of the trolley rail is 9 kg (20 lbs).
• The maximum load of the E-cylinder holder is 7.9 kg (17.5 lbs).
No maintenance is required.
5.7.12 Using the Click-in Humidifier
WARNING!
Read the chapter “Humidification” on page 24 before using the Vivo 45 with the
humidifier.
CAUTION!
The click-in humidifier and the circuit heating operates on the AC power source only.
If the AC power source fails and the internal or the external battery activates, the
click-in humidifier and the circuit heating will be turned off automatically.
NOTE
Maximum Humidifier Start-Up Period in Normal Use: 30 minutes
The click-in humidifier is intended to humidify the patient air. It is intended for non-
invasive use only. The click-in water chamber is for single patient use only. Reusing a
water chamber for a new patient might cause a risk of cross-contamination. The Vivo 45
shall not be moved with a filled water chamber installed.
5.7.12.1 Prerequisites
• The water chamber must be installed in order to access the humidifier setting on the
ventilator menu, both in clinical and home mode. If the water chamber is discon-
nected and reconnected after usage, the ventilator will remember the humidity set-
ting used.
• The click-in humidifier only operates during treatment. When the ventilator is in
standby mode, the humidification is paused.
5.7.13 Installing the Water Chamber
CAUTION!
Do not switch on the humidifier without a filled water chamber in order to avoid
burn or damage to the humidifier's electronics.
92 How to Use the Vivo 45
Vivo 45 User manual
Doc. 006149 D-3