5.6.3 Charging the Batteries
CAUTION!
Do not charge the Vivo 45 while placed in the carry bag or other types of closed or
non-ventilated spaces.
The internal and click-in batteries are automatically charged when connecting the venti-
lator to the mains supply. To ensure that the batteries are fully charged, a maintaining
charging cycle will be performed. The batteries are not charged when connecting the
ventilator to an external DC supply. While charging, the battery level will be animated.
The batteries are only charged if the internal temperatures are between 0 to 45°C (32 to
113°F). High power consuming settings in combination with high ambient temperatures
may make the battery temperature rise above 45°C (113°F).
NOTE
Charging of batteries is only started when the state of charge is below 95%
Behaviour of the ventilator while Internal Battery or Click-in Battery is
Charging
The ventilator can operate as normal while a battery is charging. The battery icon will
be animated (filling in from bottom to top) while charging.
Charging Times
Battery Charger
Time
Internal battery
Vivo 45 2 h
Click-in battery
Vivo 45
4 h
* Times are based on charging empty batteries.
5.6.4 Battery Icons
When running on battery, the battery status is indicated by the following symbols:
74 How to Use the Vivo 45
Vivo 45 User manual
Doc. 006149 D-3

Table of Contents

Other manuals for Breas Vivo 45

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Breas Vivo 45 and is the answer not in the manual?

Breas Vivo 45 Specifications

General IconGeneral
BrandBreas
ModelVivo 45
CategoryMedical Equipment
LanguageEnglish

Related product manuals