5.5.2 Insp. Pressure
The Insp. Pressure setting is used to define the airway pressure during the inspiratory
phase.
Minimum/maximum working IPAP is limited/achieved by a software control of blower
speed vs. measured pressure.
Unit Min Max Default
cmH
2
O 4 40 15
5.5.3 PEEP
The PEEP setting is used to define the airway pressure at the end of the expiratory
phase.
Unit Min
Max Default
cmH
2
O 2 20
(2)
5
(A)
2
(P)
(2)= The max setting is also limited by Insp. Pressure -2 cmH
2
O and Min Pressure
‑2 cmH
2
O.
(A)= Adult mode, (P)= Paediatric mode
5.5.4 Breath Rate
The Breath Rate setting defines the minimum number of breaths the ventilator will
deliver as long as no inspiratory trigger effort from the patient is detected. The cycles
will be ventilator-initiated breaths.
The combination of the Breath Rate and Inspiratory Time setting is limited by the I:E
ratio range 1:9.9 to 2:1.
Unit Min Max Default
1/breath
4
(A)
6
(P)
40
(A)
60
(P)
12
(A)
20
(P)
(A)= Adult mode, (P)= Paediatric mode
5.5.5 Backup Rate
The Backup Rate setting defines the minimum number of breaths the ventilator will
deliver as long as no inspiratory trigger effort from the patient is detected. The cycles
will be ventilator-initiated breaths.
The combination of the Backup Rate and Inspiratory Time setting is limited by the I:E
ratio range 1:9.9 to 2:1.
How to Use the Vivo 45 67
Doc. 006149 D-3