5.5.2 Insp. Pressure
The Insp. Pressure setting is used to define the airway pressure during the inspiratory
phase.
Minimum/maximum working IPAP is limited/achieved by a software control of blower
speed vs. measured pressure.
Unit Min Max Default
cmH
2
O 4 40 15
5.5.3 PEEP
The PEEP setting is used to define the airway pressure at the end of the expiratory
phase.
Unit Min
Max Default
cmH
2
O 2 20
(2)
5
(A)
2
(P)
(2)= The max setting is also limited by Insp. Pressure -2 cmH
2
O and Min Pressure
2 cmH
2
O.
(A)= Adult mode, (P)= Paediatric mode
5.5.4 Breath Rate
The Breath Rate setting defines the minimum number of breaths the ventilator will
deliver as long as no inspiratory trigger effort from the patient is detected. The cycles
will be ventilator-initiated breaths.
The combination of the Breath Rate and Inspiratory Time setting is limited by the I:E
ratio range 1:9.9 to 2:1.
Unit Min Max Default
1/breath
4
(A)
6
(P)
40
(A)
60
(P)
12
(A)
20
(P)
(A)= Adult mode, (P)= Paediatric mode
5.5.5 Backup Rate
The Backup Rate setting defines the minimum number of breaths the ventilator will
deliver as long as no inspiratory trigger effort from the patient is detected. The cycles
will be ventilator-initiated breaths.
The combination of the Backup Rate and Inspiratory Time setting is limited by the I:E
ratio range 1:9.9 to 2:1.
How to Use the Vivo 45 67
Doc. 006149 D-3
Vivo 45 User manual

Table of Contents

Other manuals for Breas Vivo 45

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Breas Vivo 45 and is the answer not in the manual?

Breas Vivo 45 Specifications

General IconGeneral
BrandBreas
ModelVivo 45
CategoryMedical Equipment
LanguageEnglish

Related product manuals