6.3.22 Low SpO
2
Alarm
Property Description
Alarm text
Low SpO2
Priority High
Definition A Low SpO
2
alarm will be given when the measured SpO
2
is
below the alarm limit for 30 seconds.
Possible cause
Too low flow of bleed-in oxygen.
Oxygen inlet is disconnected.
Delivered tidal volumes are too small.
Setting range 85% to 100%
Setting resolution
1%
Ventilator action The ventilator will continue treatment with the same settings.
This alarm requires a connected SpO
2
sensor.
6.3.23 High EtCO
2
Alarm
Property Description
Alarm text
High EtCO2
Priority High
Alarm condition A High EtCO
2
alarm will be given when the measured
EtCO
2
exceeds the alarm limit for 30 seconds.
Possible cause
Alarm limit is set too low.
Breath Rate too low.
Delivered Tidal Volume too low.
Excessive dead space between patient and leakage port.
Leak port occluded.
Setting range 1 to 99 mmHg
Off
Setting resolution
1 mmHg
Ventilator action The ventilator will continue treatment with the same settings.
This alarm requires a connected EtCO2 sensor
Alarms 115
Doc. 006149 D-3
Vivo 45 User manual

Table of Contents

Other manuals for Breas Vivo 45

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Breas Vivo 45 and is the answer not in the manual?

Breas Vivo 45 Specifications

General IconGeneral
BrandBreas
ModelVivo 45
CategoryMedical Equipment
LanguageEnglish

Related product manuals