Chapter 5: Setup / service program
Fresenius Medical Care 4008 S (Version V10) SM-EN 7/01.13 5-47
5.1.4.7 Setting the analog outputs of CPU 1
Explanation
Part 5
Display (field ): DAC digits (adjustable by pressing the + / – keys)
Display (field ): Analog voltage on P.C.B. LP 634 in 0.1 V
Menu item Description
A: CPU1_TEMP_SET Temperature set point
IC 9-A_2
0–5 V (X634R/C20, 0.2–10 V)
A: CPU1_DAC_DIM Dimness calibration voltage
IC 9-A_1
0–5 V (X634R/A11, 0–5 V)
A: CPU1_TEMP_ADJ Calibration voltage for temperature control
IC 9-A_20
0–5 V (X634R/C21, 0–5 V)
A: CPU1_DAC_BLL Blood leak calibration voltage
IC 9-A_19
0–5 V
(X634R/A12, 0–5 V)
A: CPU1_BIBAG_TEMP_AJ Calibration voltage for temperature NTC
(slot X12 / LP 941)
IC 18-A_2
0–5 V (X634R/A13, 0–5 V)
A: CPU1_DAC_X2 Not used
IC 18-A_1
0–5 V (X634R/C13, 0–5 V)
A: CPU1_STEUER_EP Degassing pump speed setting
IC 18-A_20
0–4.4 V (X634L/ between A, B, C 27 and A, B, C 28, 0–21 V)
A: CPU1_STEUER_FP Flow pump speed setting
IC 18-A_19
0–4.4 V (X634L/ between A, B, C 29 and A, B, C 30, 0–21 V)
Back to menu ?


O: CPU1_TEMP_SET back to menu ?
Conf
key
Conf
key
CPU1: WR ANALOG OUTP

Table of Contents

Other manuals for Fresenius Medical Care 4008 S

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fresenius Medical Care 4008 S and is the answer not in the manual?

Fresenius Medical Care 4008 S Specifications

General IconGeneral
BrandFresenius Medical Care
Model4008 S
CategoryMedical Equipment
LanguageEnglish

Related product manuals