Chapter 11: Functional description
11-8 Fresenius Medical Care 4008 S (Version V10) SM-EN 7/01.13
11.1.1.7 Venous pressure system test
Test description Check the lower limit by verifying the venous zero point.
Check the upper limit by detuning the venous pressure unit in positive
direction
(the venous line clamp remains closed while the test is in progress).
PV_DET