Chapter 10: Servicing / repair
Fresenius Medical Care 4008 S (Version V10) SM-EN 7/01.13 10-41
10.8.15.2 Microswitch
10.8.16 Filter holder for DIASAFE
®
plus
Remove the screw (2), remove the cable
holder (1), dismount the microswitch (4).
To dismount the Hall switch, unscrew the
screws from the holding block (3) and pull out
the Hall switch.
Unscrew the filter holder with 4 screws (1),
pull off the tubes.
When installing, ensure proper tubing
arrangement:
1 to valve V189
2 to CD cell CD7
3 to valve / balancing chamber