Chapter 5: Setup / service program
Fresenius Medical Care 4008 S (Version V10) SM-EN 7/01.13 5-65
5.2.2 P.C.B. LP 1631 (CPU 1) DIP switch array 2
DIP switch /
Position
Function
1
ON
OFF
CAL mode
Mode 0
Mode 1
2
ON
OFF
ON
OFF
3
ON
ON
OFF
OFF
External alarm input
Not valid
RO system
Patient bell
External alarm
4
ON
OFF
Remote control
Device with remote control
Device without remote control
5
ON
OFF
COMMCO P.C.B. LP 763
Approved
Disabled
6
ON
OFF
COMMCO
Special protocol
Default protocol
7 *
ON
OFF
Rinse section test (exclusive CDS)
Active
Not active
8
ON
OFF
Central Delivery System
Installed
Not installed
The basic position on delivery is in italics.
"No assignment" requires the OFF position.
* DIP switch 7 is no longer relevant if DIP switch 8 is set to ON.