Chapter 9: Calibration / adjustment
9-36 Fresenius Medical Care 4008 S (Version V10) SM-EN 7/01.13
9.1.40 Part 14.2: TMP-Check
9.1.41 Part 14.3: PDIAL2 press-check
Use the following steps to check flow compensation:
–Use the Arrow keys to select the Flow setting field.
– Set the flow by pressing the + / – keys
– Confirm with Audio paused key
TMP-Check
tmp: XXX mmHg
Conf
key
Use the UF pump to build up
various negative pressures
(Flow on / UF off). Compare the
values indicated on the display,
the LED display and the
external measuring instrument.
Esc
key
Note:
Observe the venous pressure!
Note:
Switch the flow on and off
again once in a while,
to maintain the operating
temperature of the pressure
transducer.
PDIAL2 press-check
PDial2: –XXX mmHg
Conf
key
Use the UF pump to build up
various negative pressures
(Flow on / UF off). Compare the
values indicated on the display,
the LED display and the
external measuring instrument.
Esc
key
Note:
Observe the venous pressure!
PDial2: –XXX mmHg
XXXX
ADC digits
Alpha
display
Flow
setting