Chapter 10: Servicing / repair
10-30 Fresenius Medical Care 4008 S (Version V10) SM-EN 7/01.13
10.8.4 Multifunction block
Pull off the cable from V43 (12) and the level
sensor (1).
Protect the other hydraulic components from
leaking fluid. Pull off the tubes and collect the
fluid (see chapter 10.8.3.1 page 10-25).
Unscrew 2 screws (10) and remove the
multifunction block.
When installing, ensure proper tubing
arrangement.
After replacement or repair, run the
degassing test and the fill program.
Legend, multi-function block
1 Level sensor 8 to V39 / orifice H89
2 to pressure measuring point B 9 to CD cell CD110
3 to degassing pump P29 10 Screws
4 to valve V91 11 to pressure measuring point D
5 to dosing point H205 12 V43, to filter F76
6 to pressure transducer S9 13 to air separation pump P97
7 to filter F74 / sensor S115